  Patented Principle of Operation      
 

Principle of Operation

The BioDAQ E2 system generates data through the use of
load cell technology. Each Peripheral Sensor Controller (PSC)
contains a load cell that reports the weight of its hopper and the
diet or water it contains once per second. As long as the weight
remains unchanged, no data is reported.

An animal initiates the recording of a feeding or drinking event by disturbing the hopper, destabilizing the reported weight. The system detects this destabilization, marks the time and date of the initiation of activity and waits for a new stable weight once the intake bout is complete. The difference between the two weights and the period of the activity are then calculated and reported along with the initiation data as a bout. BioDAQ is the only instrument that can provide this level of data collection.

U.S. Patent Numbers: 6,234,111; 6,367,418; 6,542,850; 6,748,898. U.S. and International patents pending.
BioDAQ software allows you to study the elements of eating Watch video of a rat eating in BioDAQ E2 as it records data The same electronics can be used for both food and water intake
 